High precision and working stability:

The high reliability is achieved with the installation of a CT monochromater

Advanced calculation and high automation:

New micro-processing technology is applied to strengthen the data processing.

Compatibility with PC and printer:

Built in usb interface permits bi-directional data transfer to a PC.  And a stylus printer can be connected directly to the instrument through the parallel printer port.

Easy creating and storing programmes and scans

Send your inquiry


  • Model

    TU-1900

    TU-1901

    Wavelength Range

    190-1100nm

    190-1100nm

    Spectral bandwidth

    1.8nm/1nm

    0.5/1/2/4/5nm

    Wavelength Accuracy

    ±0.3nm

    ±0.3nm

    Wavelength repeatability

    ≤0.1 nm

    ≤0.1 nm

    Photometric Range

    0~200%T -0.3-3A

    0~200%T -0.3-3A

    Photometric Accuracy

    ±0.3%T

    ±0.3%T

    Photometric Repeatability

    ≤0.15%T

    ≤0.15%T

    Stray Radiant

    ≤0.05%T  

    ≤0.05%T  

    Power Consumption

    220V  50HZ 300W

    220V  50HZ 300W

    Display

    320*240 LCD

    320*240 LCD

    Output Interface

    USB

    USB

    Size

    600*480*210

    600*480*210

    Power

    86-240AC

    86-240AC

    Net Weight

    20kg

    20kg


Phone:
+86-136 0612 1307
Telephone:
+86-(0)519-85286122/6336